| Specification:  
 
    
        
            | Humidity sensor | Capacitive (thin-film, non-condensing) |  
            | Measuring range | 0 to 100 % RH |  
            | Working range | 0 to 100 % RH; wall-mounted version 5 to 95 % RH; rod and duct version
 10 to 90 % RH; indoor version
 |  
            | Measuring accuracy | ±2.5 % RH; indoor version in humidity range from 40 to 60 % RH ±3.0 % RH; residual range
 ±2.0 % RH; wall-mounted version in humidity range of 10 to 90 % RH
 ±2.5 % RH; residual range
 ±2.0 % RH; duct version in humidity range of 40 to 60 % RH
 ±2.5 % RH; residual range
 ±2.0 % RH; rod version in humidity range from 5 to 95 % RH
 |  
            | Temperature influence | ±0.05 % RH per K; indoor and wall-mounted version (based on 23 °C) < 0.15 % RH per K; duct version (at < 10 °C, > 40 °C)
 < 0.10 % RH per K; rod version (at < 10 °C, > 40 °C)
 |  
            | Measurement medium | Air, atmospheric pressure, non-aggressive |  
            | Humidity output (active) | 4 to 20 mA 0 to 1 V (not possible with duct version)
 0 to 10 V
 |  
            | Air velocity | Min. 0.5 m/s (1.5 m/s for rod version with current output) Max. 15 m/s (wall-mounted version 10 m/s)
 |  
            | Response time | Half-period: 10 s to 1.2 min at v = 2 m/s (depending on design type and filter used, excluding indoor version)
 |  Temperature (T)
 
    
        
            | Sensing element | Pt100 or Pt1000 thin-film temperature sensor according to DIN EN 60751 (depending on version)
 |  
            | Measuring range | See order details (only scaling, observe max. admissible ambient temperature)
 |  
            | Working range | -30 to +60 °C; indoor version -30 to +80 °C; wall-mounted, duct, and rod version
 |  
            | Measuring accuracy | ±0.25 K; indoor version at voltage output ±0.4 K; indoor version at current output (10 to 40 °C)
 ±0.2 K; wall-mounted version at voltage output
 ±0.3 K; wall-mounted version at current output (10 to 40 °C)
 ±0.3 K; duct and rod version
 |  
            | Temperature influence | ±0.01 K/K (at < 10 °C, > 40 °C) |  
            | Temperature output (active) (for wire system, see wiring diagrams) | DC 0 to 1 V (not for duct version) DC 0 to 10 V
 4 to 20 mA
 |  
            | Temperature output (passive) | Pt100 (only with duct and rod version, see order details) |  Electrical data
 
    
        
            | Voltage supply Indoor and wall-mounted version 
 
 Duct version
 
 
 Rod version
 
 
 
 | DC 6 to 30 V or AC 6 to 26 V (at 0 to 1 V) DC 15 to 30 V or AC 13 to 26 V (at 0 to 10 V)
 DC 10 to 25 V (indoor version at 4 to 20 mA)
 DC 10 to 30 V (wall-mounted version at 4 to 20 mA)
 
 DC 15 to 30 V
 AC 24 V (also at 0 to 10 V)
 
 DC 12 to 30 V
 DC 15 to 30 V (at 0 to 10 V)
 DC 6 to 30 V (at 0 to 1 V)
 
 |  
            | Burden (I output) | RL(Ω)  = Voltage supply - DC 10 V /0,02 A ±50Ω |  
            | Load resistance (U output) | > 10 kΩ (at 0 to 10 V), > 2 kΩ (at 0 to 1 V) |  
            | Consumption | Typically 7 mA at voltage output, < 1 mA at 0 to 1 V (per measuring range with four-wire system)
 |  
            | Electromagnetic compatibility CE | According to EN 61326-1 and EN 61326-2-3 |  Construction
 
    
        
            | Case Indoor version
 Wall-mounted version
 Duct version
 Rod version
 
 | ABS plastic, signal white (RAL 9003)
 PC plastic
 ABS plastic with attached aluminum probe, painted
 Form J aluminum terminal head,
 probe as in duct version, optionally with 1.5 m connecting cable
 |  
            | Dimensions Indoor version
 Wall-mounted version
 Duct version
 Rod version
 | 81 × 81 × 25.7 mm
 83 × 83 × 40 mm (probe Ø 12 × 50 mm)
 248 × 120 × 48 mm (probe Ø 20 × 200 mm)
 Probe Ø 20 × 122 mm)
 |  
            | Cable entry Indoor version
 Wall-mounted version
 Duct version
 Rod version
 | Opening on the underside or rear side (flush-mounted socket assembly optional)
 via cable gland M16 × 1.5
 via cable gland M20 × 1.5
 via cable gland M16 × 1.5 (for version with form J terminal head)
 |  
            | Terminals | For conductor cross sections of up to 1.5 mm2 |  
            | Wiring diagrams | Please use the terminal connection from the operating manual included in the scope of delivery (according to version)
 |  
            | Protection type Indoor version
 Wall-mounted version
 Duct version
 Rod version
 | IP30
 IP65 (IP30 probe (standard), with PTFE IP65 sinter filter)
 IP64 (IP30 probe (standard), with PTFE IP65 sinter filter)
 IP65 (IP30 probe (standard), with PTFE IP65 sinter filter)
 |  
            | Operating temperature Indoor version
 Wall-mounted version
 Duct version
 
 Rod version
 
 | -30 to +60 °C
 -30 to +80 °C
 -40 to +80 °C at the probe head (filter area)
 -10 to +60 °C at the case
 -20 to +60 °C
 -40 to +80 °C at the probe head (filter area)
 |  
            | Installation position | For duct and rod version, install probe tube at an angle or vertically downwards, if possible. For indoor version, position vent openings downwards if possible.
 Install the measuring probe in a way that prevents ingress of water.
 Mount directly via openings in the case or using optional installation accessories.
 |  
            | Weight Indoor or wall-mounted version
 Rod or duct version
 | Approx. 200 g
 Approx. 150 to 350 g, depending on version
 |  
 |