|  | Oscilloscope specifications | Specifications with PicoConnect 442 1000 V CAT III probe | 
        
            | Input channels | 4 channels | 4 channels | 
        
            | Analog bandwidth (–3 dB) | 20 MHz with D9 to BNC adaptors 15 MHz with PicoConnect 441 probe
 | 10 MHz | 
        
            | Rise time (calculated) | 17.5 ns with D9 to BNC adaptors 23 ns with PicoConnect 441 probe
 | 35 ns | 
        
            | Bandwidth limit | 100 kHz or 1 MHz (selectable) | 100 kHz or 1 MHz (selectable) | 
        
            | Vertical resolution, 12-bit mode | 12 bits on most input ranges 11 bits on ±10 mV range
 | 12 bits | 
        
            | Vertical resolution, 14-bit mode | 14 bits on most input ranges 13 bits on ±20 mV range
 12 bits on ±10 mV range
 | 14 bits | 
        
            | Enhanced vertical resolution 12-bit mode
 | 16 bits on most input ranges 15 bits on ±10 mV range
 | 16 bits | 
        
            | Enhanced vertical resolution 14-bit mode
 | 18 bits on most input ranges 17 bits on ±20 mV range
 16 bits on ±10 mV range
 | 18 bits | 
        
            | Input type | Differential 9-pin D-subminiature, female
 | Differential 9-pin D-subminiature, female
 | 
        
            | Input characteristics | 1 MΩ ±1%, in parallel with 17.5 pF ±1 pF (each differential input to ground). < 1 pF difference between ranges.
 | 16.7 MΩ ±1%, in parallel with 9.3 pF ±1 pF (each differential input to ground) | 
        
            | Input coupling | AC/DC | AC/DC | 
        
            | Input sensitivity (10 vertical divisions)
 | 2 mV/div to 10 V/div | ±0.5 V/div to ±200 V/div | 
        
            | Input ranges (full scale) | ±10 mV, ±20 mV, ±50 mV, ±100 mV, ±200 mV, ±500 mV, ±1 V, ±2 V, ±5 V, ±10 V, ±20 V, ±50 V
 | ±2.5 V, ±5 V, ±12.5 V, ±25 V, ±50 V, ±125 V, ±250 V, ±500 V, ±1000 V | 
        
            | Input common mode range | ±5 V on ±10 mV to ±500 mV ranges ±50 V on ±1 V to ±50 V ranges
 | ±125 V on ±2.5 V to ±12.5 V ranges ±1000 V on ±25 V to ±1000 V ranges
 | 
        
            | DC accuracy (10 kHz) | ±1% of full scale ±500 µV | ±3% of full scale ±12.5 mV | 
        
            | Analog offset range | ±250 mV on ±10 mV to ±500 mV ranges ±2.5 V on ±1 V to ±5 V ranges
 ±25 V on ±10 V to ±50 V ranges
 | ±6.25 V on ±2.5 V to ±12.5 V ranges±62.5 V on ±25 V to ±125 V ranges
 ±625 V on ±250 V to ±1000 V ranges
 | 
        
            | Analog offset accuracy | 1% of offset setting in addition to basic DC accuracy | 1% of offset setting in addition to basic DC accuracy | 
        
            | Overvoltage protection | ±100 V DC + AC peak (any differential input to ground) ±100 V DC + AC peak (between differential inputs)
 | CAT III 1000V 
 |