            | Oscilloscope — vertical |  
            | Model | PicoScope 4424 | PicoScope 4224 | PicoScope 4224 IEPE |  
            | Passive probe mode | IEPE interface mode |  
            | Input channels | 4 | 2 |  
            | Bandwidth | DC to 20 MHz | 1.6 Hz to 20 MHz |  
            | (10 MHz on ±50 mV range) |  
            | Rise time (10% to 90%, calculated) | 17.5 ns (35 ns on ±50 mV range) |  
            | Input type | Single-ended, BNC connector |  
            | Vertical resolution | 12 bits |  
            | Enhanced vertical resolution | 16 bits |  
            | Input sensitivity | 10 mV/div to 20 V/div | 10 mV/div to 4 V/div |  
            | Input ranges (full scale) | ±50 mV to ±100 V in 11 ranges | ±50 mV to ±20 V in 9 ranges |  
            | Input coupling | Software-selectable AC/DC |  
            | Input characteristics | 1 MΩ || 22 pF | 1 MΩ || 1 nF |  
            | DC accuracy | ±1% |  
            | Overvoltage protection | ±200 V (DC + AC peak) | ±100 V (DC + AC peak) |  
    
    
        
            | Oscilloscope — horizontal |  
            | Maximum sampling rate (single shot) | 1 or 2 channels: 80 MS/s* 3 or 4 channels: 20 MS/s
 | 80 MS/s |  
            | * To achieve the stated sampling rate with two channels, enable one channel from A or B and one from C or D. |  
            | Maximum sampling rate (continuous streaming mode)
 | 10 MS/s (Record length limited to 20 MS in PicoScope, unlimited when using the supplied SDK) |  
            | Timebase ranges | 100 ns/div to 5000 s/div |  
            | Buffer memory | 32 M  samples shared between active channels |  
            | Maximum buffer segments | 10,000 |  
            | Timebase accuracy | 50 ppm with ≤10 ps jitter |  
    
    
        
            | Triggers |  
            | Source | Any input channel |  
            | Trigger modes | Free run, auto, repeat, single, rapid (segmented memory) |  
            | Trigger types | Rising edge, falling edge, edge with hysteresis, pulse width, runt pulse, dropout, logic, windowed |  
            | Trigger level | Adjustable over whole of selected voltage range |  
            | Trigger sensitivity | Digital triggering provides 1 LSB accuracy up to full bandwidth. |  
            | Maximum pre-trigger capture | 10 x “per division” timebase setting |  
            | Maximum post-trigger delay | 50 s |  |