| Specifications | 
        
        
            | Product start-up time | 
            5 seconds minimum | 
        
        
            | Measurement period | 
            5–180 seconds | 
        
        
            | Reporting interval | 
            0–3600 seconds via RS232 communication | 
        
        
            | Principle of operation | 
            Laser Diode optical detection of actual particulates | 
        
        
            | International codes | 
            ISO 7 – 22, NAS 0 – 12 | 
        
        
            | Calibration | 
            By recognised online methods confirmed by the relevant ISO procedures: MTD – via a certified primary ISO 11171 automatic particle detector using ISO 11943 principles, with particle distribution reporting to ISO 4406:1996 | 
        
        
            | Recalibration | 
            Contact Parker Hannifin | 
        
        
            | Working pressure | 
            2–420 bar (30–6000 PSI) | 
        
        
            | Flow range through icountPDR | 
            Note : Flow may be bi-directional 40–140 ml/min (optimum flow 60 ml/min) (0.01 – 0.04 USGPM (optimum flow 0.016 USGPM)) | 
        
        
            | Online flow range via System 20 sensors | 
            Size 0 = 6 to 25 l/min (2–7 USGPM) Size 1 = 24 to 100 l/min (6–26 USGPM) Size 2 = 170 to 380 l/min (45–100 USGPM) | 
        
        
            | Ambient storage temperature | 
            –40°C to +80°C (–40°F to +176°F) | 
        
        
            | Environment operating temperature | 
            -30°C to +60°C (–22°F to 140°F) | 
        
        
            | Fluid operating temperature | 
            +5°C to +80°C (+41°F to 176°F) | 
        
        
            | Computer compatibility | 
            Parker recommends the use of a 9-way D-type connector. This can be connected to a USB port using a USB-serial adaptor. Note that these connectors/adaptors are NOT supplied with icountPDR units: contact Parker Hannifin for advice. | 
        
        
            | Moisture sensor calibration | 
            ±5% RH (over compensated temperature range of +10ºC to +80ºC; +50ºF to +176ºF) | 
        
        
            | Operating humidity range | 
            5% RH to 100% RH | 
        
        
            | Moisture sensor stability | 
            ±0.2% RH typical at 50% RH in one year | 
        
        
            | Power requirement | 
            Regulated 9–40Vdc | 
        
        
            | Current rating | 
            Typically 120mA | 
        
        
            | Certification | 
            IP69K rating | 
        
        
            | Analogue output options (specified when ordering) | 
        
        
            | Variable current | 
            4–20mA | 
        
        
            | Variable voltage | 
            0–5Vdc, 0–3Vdc (user selectable) | 
        
        
            | CAN-bus | 
            to SAE J1939 (e.g. Parker IQAN) | 
        
        
            | Moisture sensor | 
            Linear scale within the range 5% RH to 100% RH |