Pond Engineering designed this stirred liquid bath system to provide user-friendly calibrations over the range of -80°C to +300°C in a stable, uniform environment. Three optional configurations allow the user to tailor the system to meet the needs of their particular calibration workload. The base system is comprised of two well modules positioned side by side in a single cabinet and covers the range of temperatures from -80°C to + 300°C. The low-temperature well is capable of operating between -80°C and +70°C and will typically utilize a water/glycol solution and methanol bath fluids to cover that range. The high-temperature well operates between +50°C and +300°C using silicone oils of various viscosities.
An interactive user interface on a sloping front panel provides convenient system control in a stand-alone configuration, and can be configured to display in either °C or °F. The system setpoint can be set directly (no additional variables or verniers to tweak) using the front panel interface. Optional remote interface allow users to integrate the K61 into an automated calibration system, dramatically increasing throughput and reducing the cost of calibrations.