            | Model | TGI-250kN | TGI-100kN | TGI-50kN | TGI-10kN | TGI-5kN | TGI-1kN | TGI-500N | TGI-200N | 
        
            | Maximum test force | 250 kN | 100 kN | 50 kN | 10 kN | 5 kN | 1 kN | 500 N | 200 N | 
        
            | Test force measure -ment range *1) | 0.1N~ 250kN
 | 0.1N~ 100kN
 | 0.1N~ 50kN
 | 0.1N~ 10kN
 | 0.1N~ 5kN
 | 0.1N~ 1kN
 | 0.1N~ 500N
 | 0.1N~ 200N
 | 
        
            | Test space | Tensile / compression testing beneath the movable crosshead. (A model is available allowing tensile testing above the movable crosshead (Optional).)
 | 
        
            | Loading method | Constant-speed strain loading with highly accurate AC servo motor and backlash-free precise ball screw. | 
        
            | Cross head speed / permissible test force | Within the max. test force over the entire speed ranges. | 
        
            | Test force range | 2 methods: AUTO: (accuracy guranteed for 1/1 to 1/250 of load cell capacity)
 100: (accuracy guranteed for 1/100 to 1/1 000 of load cell capacity)
 | 
        
            | Test force display | 5-digits digital display ±99,999 AUTO range: Resolution 1/5000, test force reduction increase/decrease, Full scale display, and automatic switching of decimal point position and display unit.
 100 range: Resolution 1/10000
 * Test force display varies depending on the load cell provided.
 | 
        
            | Test force measuring accuracy (Guaranteed accuracy range) | Standard accuracy : ±1 % of indicated value (1/1 to 1/1000 of load cell capacity)
 High accuracy : ±0.5% of indicated valeu (Optional)
 (1/1 to 1/1000 of load cell capacity)
 * Test machine must be in location free of vibration
 | 
        
            | Applicable standard | JIS B 7721 Class 1 or Class 0.5 and ISO7500-1 Class 1 or Class 0.5 | 
        
            | Effective test width (Note 2) (mm) | 600 | 600 | 420 | 430 | 
        
            | Cross head stroke (mm) | 1 550 | 1 350 | 1 015 | 1 090 | 
        
            | Maximum mounting space (mm) | 1 690 | 1 555 | 1 165 | 1 160 | 
        
            | Tensile stroke (Note 3) (mm) | 620 | 850 | 470 | 710 | 820 | 870 | 890 | 910 | 
        
            | Crosshead speed | 0.000 5 to 500 mm/min 0.1 to 500 mm/mincan be set in 0.1 mm/min incliments. Below this, can be set in 9 steps: 0.05, 0.03, 0.02, 0.01, 0.005, 0.003, 0.002, 0.001, and 0.0005.
 | 0.000 5 to 1 000 mm/min 0.1 to 1 000 mm/min; can be set in
 0.1 mm/min increments. Below this, can be set in 9 steps: 0.05, 0.03, 0.02, 0.01, 0.005, 0.003, 0.002, 0.001, and 0.0005.
 | 
        
            | Crosshead speed accuracy | ±0.1 % over entire speed range | 
        
            | Return speed | Can be set as required between 0.1 mm/min and 750 mm/min | Can be set as required between 0.1 mm/min and 1 500 mm/min | 
        
            | Speed setting method | Manual: Numeric input via control panel When using data processing software: Automatically set in accordance with test conditions created.
 | 
        
            | Manual speed | 3 speeds (HIGH/MIDDLE/LOW); each freely settable | 
        
            | Test mode | Single and cycle (Combination of test force and position) Optional: Creep, Automatic loading
 | 
        
            | Test force display | 5 digit digital display ±99 999 Auto range: Resolution 1/5 000, test force reduction increase/decrease, full scale display, and automatic switching of decimal point position and display units
 100 range: Resolution 1/10 000 * Test force display varies depending on the load cell provided.
 | 
        
            | Movable crosshead position display | 6 1/2-digit digital display ±1999.999 | 
        
            | Standard function | 
                
                    
                        | Load cell model & capacity | : | Auto adjust |  
                        | Load cell span | : | Auto set for tension & compression |  
                        | Zero point | : | One push |  
                        | Position reset | : | One push |  
                        | Sample rupture detection | : | Selectable from stop, reverse and ignore |  
                        | Rupture sensitivity | : | Set as desired to 0.04 % to 99.00 % of test force |  
                        | Test condition registration function | : | 10 conditions |  
                        | Test force & position analog output | : | DC0 ~ DC5 V (with scaling function) |  
                        | Interface | : | Conforms to USB interface 2.0/1.1 standard (B- type connector) |  | 
        
            | Options | 
                
                    
                        | Chuck adapter | : | Lower chuck mounting adapter for tensile testing |  
                        | Load cell | : | for tension / compression type (with universal joint, connect pin and compression plate
 ----------
 for tensile type
 (with universal joint and connecting pin)
 ----------
 for compression type
 (with for compression plate)
 |  
                        | Testing jig | : | Former jig is attachable. |  
                        | Displacement measuring function | : | Strain input type |  
                        | Test machine stand |  
                        | Data processing software | : | Standard (Tensile / compression / bending), Peeling, Cycle (One way, Both way), Stress relaxation
 |  | 
        
            | External dimensions (WxHxD)(mm) | 1420 x 2406 x 894 | 1221 x 2192 x 720 | 900 x 1507 x 490 | 700 x 1583 x 523 | 700 x 1 583 x 513 (without level adjuster)
 | 
        
            | Approximate weight (kg) | 1 200 | 900 | 310 | 120 | 
        
            | Power supply voltage, Frequency | 3-phase AC200 V to 230 V+10 % to 15 %, 50/60 Hz | Single phase AC100 V to 115 V+10% to -15%, 50/60 Hz | 
        
            | Power consumption (W) | 7.5 k | 2.9 k | 1.8 k | 750 | 400 | 
        
            | Operating temperature / humidity range | Temperature 5 ℃ ~ 40 ℃, Humidity 20 % ~ 80 %RH (with no condensation)
 |