Máy đo độ cứng siêu âm UCI NewSonic SonoDur2
 
 
 Measuring Specification
    
        
            | 
             Measuring principle 
             | 
            
             UCI Method, corresponds to DIN 50159, ASTM A1038 
             | 
         
        
            | 
             Test indenter 
             | 
            
             Vickers diamond 136° 
             | 
         
        
            | 
             Test loads 
            Newton scale (1kgf = 9.81 N) 
             | 
            
             Motor probes: 1N (0.1 kgf), 3N (0.3kgf) and 8.6 N (0.8 kgf) 
             | 
         
        
            | 
             Handheld Probes: 10N (1 kgf), 49N (5kgf), 98N (10kgf) 
             | 
         
        
            | 
             (Other test loads on request) 
             | 
         
        
            | 
             Hardness scales and range 
              
            Note: 
            Conversions are acc. to latest ASTM E140, EN ISO 18265, and DIN 50150. Conversions into tensile strength: 98N (10kgf) test load only. 
             | 
            
             Vickers HV 10 – 9999 
             | 
         
        
            | 
             Brinell HB 76 – 618 
             | 
         
        
            | 
             Knoop HK 87 – 920 (ASTM only) 
             | 
         
        
            | 
             Rockwell HRB 41 – 105 
             | 
         
        
            | 
             Rockwell HRF 82,6 – 115,1 
             | 
         
        
            | 
             Rockwell HRC 20,3 – 68 
             | 
         
        
            | 
             Rockwell HRA 60,7 – 85,6 
             | 
         
        
            | 
             Rockwell HRD 40,3 - 76,9 (EN ISO 18265 only) 
             | 
         
        
            | 
             HR45N 19,9 – 75,4 
             | 
         
        
            | 
             Tensile Strenght MPa (N/mm2) 255 – 2180 (EN ISO 18265 only) 
             | 
         
        
            | 
             Measurement uncertainty 
             | 
            
             < 3% of the average out of 5 measurements relative to the plate value 
             | 
         
        
            | 
             Relative repeatability 
             | 
            
             < 3% (range relative to the average out of 5 measurements on reference block 300HV using motor probes 3N, 8.6N) 
             | 
         
    
 
 
  
                             |