| Measuring range: |
|
| Frost/Dew Point |
-20 ...+70 °C |
| % RH |
5 ...100% RH |
| Temperature |
-50...+100 °C |
| Accuracy: |
|
| Frost/Dew point |
SH2: ± 0.1 °C
RP2: ± 0.15 °C |
| Temperature |
± 0.07 °C |
| Reproducibility: |
|
| Frost/Dew Point |
± 0.07 °C |
| Temperature |
± 0.05 °C |
| Standard Features |
|
| Temperature Probe |
SH2: Ø2 x 100 mm PRT, 0.5 and 3 m cables
RP2: Ø3 x 30 mm PRT, 0.5 m cable |
| Digital I/O |
RS-232 |
| Display |
5.7" LCD with color touch screen |
| Mirror Temperature Sensor |
Platinum Resistance Thermometer (Pt100) |
| Gas Connections |
SH2: 6 mm or 1/4" Swagelok
RP2: Not Applicable |
| Transport Case |
Custom fit foam lined Peli 1550 |
| Power Cable |
2.5 m, multiple countries available |
| Operating Instructions |
English |
| Calibration Certificate |
Factory calibration: 5 points FP/DP, 3 points temperature |
| Optional Features |
|
| Internal Barometric Pressure Sensor |
0.1% or 0.01% accuracy, 700...1200 mbar |
| Analog Outputs |
Two user-programmable, -10...+10 V and 4...20 mA |
| Calibration Upgrade |
Upgrade to ISO 17025 calibration |
| Power Supply |
100-120V VAC / 200-240 VAC, 50/60 Hz, 100 Watt (auto switching) |
| Operating Conditions |
|
| Instrument |
0...40 °C, 90% RH non-condensing |
| Measuring Head |
-50...80 °C, 99% RH non-condensing |
| Storage Conditions |
-20...50 °C |
| Weight & Dimensions |
|
| Dimensions: |
Instrument: W310 x H155 x D265 mm
In Transport Case: W510 x H220 x D450 mm |
| Weight: |
Instrument: 5 kg
In Transport Case: 12 kg |
| Protection |
Instrument: IP20
In Transport Case: IP65 |