| Humidity | 
        
            | Generation Range | 10-90% RH | 
        
            | Accuracy Control Element
 | ≤±1% RH (10–70% RH) ≤±1.5% RH (70–90% RH)
 | 
        
            | Stability | ±0.2% RH (20–80% RH) | 
        
            | Temperature | 
        
            | Generation Range | +10 to +50°C (Lowest T set point = 10°C below ambient)
 | 
        
            | Accuracy | ±0.1°C | 
        
            | Stability | ±0.1°C | 
        
            | Chamber | 
        
            | Ramp Rate from +20 to +40°C
 +40 to +20°C
 | 1.5°C/minute
 0.7°C minute
 | 
        
            | Control Element | Removable relative humidity sensor | 
        
            | General | 
        
            | Probe Ports | up to 5 – sensor body diameters 5 to 25mm accommodated by port adapters | 
        
            | Chamber Volume | 2000cm³ | 
        
            | Chamber Dimensions | 105 x 105 x 160mm (h x w x d) | 
        
            | Instrument Dimensions | 290 x 520 x 420mm (h x w x d) | 
        
            | Set Point Resolution | 0.1 for humidity and temperature | 
        
            | Displays | 3 digit LED, 10mm characters | 
        
            | Supply | 85 to 264 V AC, 47/63 Hz, 150 VA | 
        
            | Weight | 20kg |