| Sampling Method & Range | Diffusion, 0-25% O2, 0-10,000ppm CO2 | 
        
            | Accuracy | ± 1% of full scale | 
        
            | Operating Temperature | -0 to + 55 C | 
        
            | Display | ¾” backlit LCD digital display | 
        
            | Sensor Type | Long life zirconium oxide sensor for O2 and NDIR for CO2 | 
        
            | Alarm indicators | Built-in Audible 90dB horn, Orange LED alarm indicator for CO2, Red LED alarm indicator for O2 | 
        
            | Signal Outputs | One user adjustable Alarm relay for Oxygen, One user adjustable Alarm relay for CO2. Relays are rated 2 amp 24VDC / 24VAC | 
        
            | Analog Output | Analog 4-20mA output, selectable for either Oxygen or CO2. Optional 0-5 VDC output for CO2. | 
        
            | Power Requirements | 24VDC 350mA | 
        
            | Dimensions | 6.5 (W) x 4.5 (H) x 3.25 (D) inches; (165.1 mm x 114.3 mm x 82.5 mm) | 
        
            | Weight | 1.6 lbs. (.8 kg) | 
        
            | Enclosure | Polycarbonate general purpose | 
        
            | Required calibration | None for Oxygen. None for CO2 when Auto Cal function is activated. |