| Bridge power supply |
DC5 V±0.25 V within 60 mA
(changeable to DC2.5 V) |
| Input range |
F.S. setting is available with the input range from -3.1 mV/V to 3.1 mV/V |
| Output range |
DC±5 V Load resistance is 5 kΩor more.
(F.S. setting is available by function) |
| Output times |
100 times/s(Changeable to 10 times/s or 20 times/s or 50 times/s) |
| Output resolution |
1/12 000 and above |
| Zero adjustment range |
-2.5 mV/V to 2.5 mV/V |
| Non-linearity |
Display 0.02 %F.S.、Output 0.04%F.S. |
| Temperature efficient |
Zero ± 0.2 μV/℃、Sensitivity ± 0.001 5 %F.S./℃
(When the calibration is made at 0.2 μV/d or more of the input sensitivity.) |
| Input noice |
± 0.2 μVp−p or less (At the default setting of all filter.) |
| Digital low pass filter |
0.1 Hz to 10 Hz (digital, adjustable in 0.1 Hz increments) |
| A/D sampling speed |
100 times/s (Changeable to 10, 50, 100 times/s.) |
| Display range |
-99 999 to 99 999 |
| Display times |
4 times/s (Changeable to 20 times/s or 50 times/s or 100 times/s ) |
| Decimal point display |
Changeable to No display, 10¹, 10², 10³, 10⁴ |
| External control function |
4 points ZERO、PEAK/TRACK/A/Z、HOLD、RESET/A/Z OFF |
| Comparator function |
Set value :-99 999 to 99 999
Set quantity :2 points of S1 and S2
Conversion times of comparator :Synchronous with A/D sampling |
| Operating temp. /humidity range |
Temperature -10 ℃ to 50 ℃, Humidity 85 %RH or less (Non condensing) |
| Power supply voltage |
100 VAC to 240 VAC 50/60 Hz
(Permissible variable range 85 VAC to 264 VAC) |
| Power consumption |
Max. approx. 9.2 VA (with option, at 230 V AC), max. approx. 5.6 VA (with option, at 100 V AC) |
| Outline dimensions (WxHxD) |
96 mm×48 mm×138 mm |
| Weight |
Approx. 0.35 kg (Options, including mounting kit) |
| Protection code |
IP65 equivalent (Only for the fron panel when the attached panel mounting gascket is installed. |