| Typical measuring time | 3–5 minutes | 
        
            | Maximum allowed volume of packages to be handled by PFD/PFD Plus | Can: 0.5 L Glass bottle: 1 L
 PET bottle: 3 L
 | 
        
            | Sample temperature | 15–25 °C (59–77 °F) | 
        
            | Sample volume per measurement | 120–150 mL (approx. 20 mL per additional measuring module)
 | 
        
            | Compressed gas:necessary quality of air if used as compressed gas supply
 | Class 5 from ISO 8573-1 max. particle size: 40 μm
 max. pressure dew point: +7 °C (44.6 °F)
 max. oil content: 25 mg/m³
 | 
        
            | Compressed gas: necessary quality of nitrogen supply if Option O2 is installed:
 | Purity ≥ 99.999 % Class 5.0
 max. particle size: 40 μm
 max. pressure dew point: +7 °C (44.6 °F)
 | 
        
            | Operating pressure for PBA system | 6 ±0.5 bar relative (87 ±7 psi) | 
        
            | Dimensions (L x W x H) of the PBA system | 495 x 700 x 670 mm (19.5 x 27.6 x 26.4 in) | 
        
            | Voltage | AC 100 to 240 V, 50/60 Hz | 
        
            | Power consumption | 190 VA (incl. external measuring modules) | 
        
            | Power inlet | according to IEC/EN 60320-1/C14, protection class I | 
        
            | Fuses | Glass tube fuses 5 x 20 mm; IEC60127-2; AC 250 V; T 5 AH | 
        
            | Environmental conditions (EN 61010) | Indoor use only | 
        
            | Ambient temperature | 15–35 °C (59–95 °F) | 
        
            | Air humidity | 10–90 % relative humidity, non-condensing | 
        
            | Pollution degree | 2 | 
        
            | Overvoltage category | II | 
        
            | Altitude | max. 3000 m (9800 ft) |