Standards the instrument complies with
ISO 48; ASTM D 1415; DIN 53 519
 
Unit of measure
IRHD-M (micro)
 
Resolution
0.1 irhd point
 
Instrument Control
With Gibitre-Hardness software
 
Test modality
Serial automatic testing of the parts placed across the test line of the sample holding disk
 
Calculated Results
• IRHD Hardness • Thickness of the sample • Angle coefficient of hardness relaxation curve • Hysteresis (sample return after set time from load removal) • Correction of hardness according to the thickness of the sample (according to WIP ISO 48:2010)
 
Data analysis
Mean, std. Dev., min, max, Cp, Cpk of test results. X-Chart and Gaussian distribution
 
Graphs
Rubber Relaxation curve (hardness versus test) time in linear and logarithmic axes
 
Tolerance verification
Comparison of test results with the tolerance limits set for the product
 
Results storage
Saving of results and curves in standard database
 
Laser Device
Class 2 laser sensor Resolution: 0.002 mm
 
Sample thickness
Between 1 and 20 mm
 
Calibration
Electronic calibration Report with traceability to primary references ACCREDIA calibration Certificate (optional)
 
Personal computer
Minimum Configuration: Intel Core I3 2 GB RAM Compatible operating Systems: 7 and 8 (64 bits) Connection to the instrument via USB cable (included)
 
Software usage Languages
Italian, English, French, Spanish, German, Portuguese, Russian, Chinese, Japanese, Turkish, Polish
 
Power supply
110÷240 VAC ±10%, 50÷60 Hz ±3, 4 A, Single Phase 40 W
 
Dimensions
(W x D x H) 470 x 350 x 600 mm