| Type |
Portable or panel-mounted |
| Display system |
4-digit digital display (O2 concentration)
Bar graph display (atm) |
| Measurement range |
0-1 / 10 / 100 / 1000 ppm / 1 / 10 / 100 vol%O2
10-20-10-0atm |
| Repeatability |
Within ±0.5 % of full scale (at or above 1 % range)
Within ±1 % of full scale (below 1 % range)
(Not guaranteed for the 0-10ppm range or below) |
| Sampling system |
Continuous suction |
| Sample gas rate |
1000-2000 mL/min |
| Sample gas conditions |
Pressure: Pressurized to about 0.03MPa
Flow rate: 1000-2000 mL/min
Humidity: Dew point lower than ambient temperature
Other: Must not contain flammable components, halogen, silica, corrosive components or water droplets. |
| Reference gas |
Air |
| Recorder output |
External output: DC 0-10 V
DC4-20mA
Range marker: BCD, 3 contact outputs (Contact capacity: DC 24 V, 0.3 A) |
| Alarm output |
Upper/Lower limit alarm |
| Ambient temperature |
0-40 °C |
| Warmup time |
Approx. 20 min |
| Dimensions |
210(W) × 134(H) × 376(D) mm |
| Weight |
Approx. 7 kg |
| Panel cut dimensions |
225(W) × 135(H) |
| Gas connection |
INLET (Sample gas inlet): Rc1/4
BYPASS OUT (Bypass outlet): Rc1/4
OUTLET (Sample gas outlet): Rc1/4
|
| Power supply |
AC 100 V 50/60 Hz |
| Max. power consumption |
Less than 150 VA |