| Type |
Portable or panel-mounted |
| Display system |
O2 conc.; LED Digital, 6 Digitsl |
| Measurement range |
0.1ppm?100vol%O2
(Oxygen Concentration (less than 1ppm) and 0-1ppm or less is out of warranty.) |
| Response speed |
In 30 sec. after changing sampling gas from Inlet. (90% response) |
| Repeatability |
Within ±1 % of full scale (Not guaranteed for 1ppm and the 0-1 ppm display range or below) |
| Sampling system |
Continuos Suction |
| Sample gas rate |
200±50 ml/min |
| Sample gas conditions |
Flow Rate; 1000~2000 ml/min
Pressure; 0.03MPa or less
Humidity; Dew point shall be lower than ambient temperature.
No Flammable Gas, No Halogen Gas, Silica, Corrosive Gas nor Vapor shall be contained. |
| Recorder output |
Output signal: DC 0-10 V, insulation/load resistance 500 Ω max. |
| Communication |
Conformed to RS-232C |
| Alarm output |
Device Error Output
Concentration Error |
| Self Diagnosis |
Warm-up Error, CPU Error,Instruments Over Heat Error,and other 11type |
| Automatic refresh function |
Apply current to sensor automatically in detecting the deterioration of sensor during air calibration |
| Ambient temperature |
0-40 °C |
| Warmup time |
In 20 min. |
| Dimensions |
288(W) × 144(H) × 400(D)mm (Excluding protrusion) |
| Weight |
Approx. 10 kg |
| Gas connection |
INLET (Sample Gas); Rc 1/8
BYPASS OUT ; Rc 1/8
OUTLET (Sample Gas); Rc 1/4 |
| Power Source |
Voltage; AC 100V±10 V?Frequency; 50/60 Hz |