| System Specifications | 
            Value | 
        
        
            | Pump | 
        
        
            Flow rate range 
            Analytical gradient 
            Loading gradient (-TE model) | 
              
            5-50 µL/min or 20-200 µL/min (depends on model) 
            20-200 µL/min | 
        
        
            | Maximum pressure 10,000 psi | 
            Maximum pressure 10,000 psi | 
        
        
            | Retention time reproducibility | 
            < 0.5% RSD (@20 µL/min; 5-50 µL/min model) | 
        
        
            | Gradient delay volume | 
            < 3 µL | 
        
        
            | Autosampler | 
        
        
            | Sample capacity | 
            6 micro titer plates (96 or 384 well) 
            6 vial adapters for 54 x 2 ml vial | 
        
        
            | Injection volume range | 
            2-50 µL | 
        
        
            | Injection volume precision | 
            < 1% RSD full loop 
            < 2% RSD partial loop | 
        
        
            | Carryover | 
            < 0.005% (benzophenone) | 
        
        
            | Injection/auxiliary valve | 
            6 port 
            6-32 thread size (1/32") 
            port-to-port volume < 60 nL | 
        
        
            | Oven | 
        
        
            | Temperature range | 
            +5°C to 60°C | 
        
        
            | Inside dimensions (L x W x D) | 
            8" x 3/4" x 5/8" (20 x 1.9 x 1.6 cm) | 
        
        
            | System | 
        
        
            | Instrument control | 
            Controlled by SCIEX Analyst® 1.6.2 Software or higher, and TF 1.7 software or higher | 
        
        
            | Dimensions (L x W x D) | 
            33.5" x 22.8" x 27.6" (85 cm x 58 cm x 70 cm) |