| Specifications |
| Infrared temperature range |
-40 °C to 800 °C
(-40 °F to 1472 °F) |
| Infrared accuracy |
< 0 °C (32 °F): ± (1.0 °C (± 2.0 °F) + 0.1°/1 °C or °F);
> 0 °C (32 °F): ± 1 % or ± 1.0 °C (± 2.0 °F), whichever is greater |
| Display resolution |
0.1 °C / 0.1 °F |
| Infrared spectral response |
8 µm to 14 µm |
| Infrared response time |
< 500 ms |
| Thermocouple type-k input temperature range |
-270 °C to 1372 °C (-454 °F to 2501 °F) |
| Thermocouple type-k input accuracy |
<-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is greater <-40 °F: ±(2 °F + 0.2 °/1 °F)
≥-40 °F: ±1 % or 2 °F, whichever is greater |
| D:S (distance to measurement spot size) |
50:1 |
| Laser sighting |
Single-point laser |
| Minimum spot size |
19 mm (0.75 in) |
| Emissivity adjustment |
By built-in table of common materials or digitally adjustable from 0.10 to 1.00 by 0.01 |
| Data storage |
99 points |
| Hi/Low alarms |
Audible and two-color visual |
| Min/Max/Avg/Dif |
Yes |
| Display |
Dot matrix with function menus |
| Backlight |
Two levels, normal and extra bright for darker environments |
| Trigger lock |
Yes |
| Switchable celsius and fahrenheit |
Yes |