General Specifications
|
| Temperature Range |
50 °C to 680 °C |
| Ambient Operating Range |
5 °C to 45 °C |
| Stability |
±0.03 °C to 300 °C
±0.05 °C above 300 °C |
| Vertical Gradient |
Top and bottom zones adjustable by offset |
| Plateau Duration |
6–10 hours typical |
| Resolution |
0.01 ° |
| Display Scale |
°C or °F, switchable |
| Immersion Depth |
229 mm (9 in) |
| Stabilization Time |
15 minutes nominal |
| Preheat Wells |
2 |
| Fault Protection |
Sensor burnout and short protection, over-temperature thermal cutout |
| Display Accuracy |
±0.2 °C to 300 °C
±0.3 °C to 450 °C
±0.5 °C to 680 °C |
| Comparison Block |
Two multi-hole blocks, blanks, and custom blocks available |
| Well-to-Well Gradient (in comparison block) |
±0.02 °C |
| Heating Time |
1.25 hrs. from 25 °C to 680 °C |
| Cooling Time |
10.5 hrs. from 680 °C to 100 °C |
| Comm. |
RS-232 included |
| Power Requirements |
115 VAC (±10 %), 60 Hz, 11 A, or
230 VAC (±10 %), 50 Hz, 6 A, specify, 1200 W |
| Exterior Dimensions (HxWxD) |
489 x 222 x 260 mm
(19.25 x 8.75 x 10.25 in) |
| Weight |
20.5 kg (45 lb.) with block |