Specifications
|
| Temperature Range |
–10 °C to 125 °C |
| Ambient Operating Range |
5 °C to 45 °C |
| Stability |
±0.02 °C |
| Vertical Gradient |
±0.05 °C over 100 mm at 0 °C |
| Plateau Duration |
6–10 hours, typical |
| Resolution |
0.01 ° (0.001 ° in program mode) |
| Display Scale |
°C or °F, switchable |
| Immersion Depth |
171 mm (6.75 in) in optional comparison block |
| Stabilization Time |
15 minutes nominal |
| Preheat Wells |
3 wells (for 3.18, 6.35, or 7.01 mm probes [0.125, 0.25, 0.276 in]) |
| Fault Protection |
Adjustable software cutout using control probe; separate circuit thermocouple cutout for maximum instrument temperature |
| Display Accuracy |
±0.25 °C |
| Comparison Block |
Three multi-hole blocks, blanks, and custom blocks available |
Well-to-Well Gradient
(in comparison block) |
±0.02 °C
|
| Heating Time |
Ambient to 100 °C: 45 min. |
| Cooling Time |
Ambient to –5 °C: 25 min. |
| Comm. |
RS-232 included |
Power Requirements
|
115 VAC (±10 %), 60 Hz, 1.5 A, or
230 VAC (±10 %), 50 Hz, 0.75 A, 170 W |
| Exterior Dimensions (HxWxD) |
489 x 222 x 260 mm (19.25 x 8.75 x 10.25 in)
|
| Weight |
7 kg (15.5 lb.) with block |