| DC voltage, output |
| Range [1] |
Absolute Uncertainty, ±
(ppm of output + μV), 1-yr |
Resolution |
| 0 to 100 mV |
30 |
3 |
1 μV |
| 0 to 1 V |
30 |
10 |
10 μV |
| 0 to 10 V |
30 |
100 |
100 μV |
| 0 to 100 V |
30 |
1 mV |
1 mV |
| TC Output and Input |
| –10 mV to 75 mV |
30 |
2 μV |
1 μV |
[1] All outputs are positive only unless otherwise noted
| DC voltage, isolated input |
| Range |
Absolute Uncertainty, ±
(ppm of reading + mV),
1-yr |
Resolution |
| 0 to 10 V |
50 |
0.2 |
100 μV |
| 10 V to 100 V |
50 |
2.0 |
1 mV |
| DC Current, output |
| Range[1] |
Absolute Uncertainty,
± (ppm of reading + μA),
1-yr |
Resolution |
| 0 to 100 mA |
50 |
1 |
1 μA |
|