| Operating limits |
| Output ranges |
2 A, 20 A, 120 A (100 A dc) |
| Input current range |
200 mA, 200 mA, 120 mA (100 mA dc) |
| Current gain |
10, 100, 1000 |
| Input voltage range |
2 V, 2 V, 1.2 V (1.0 V dc) |
| Transconductance |
1,10,100 Siemens |
| Frequency |
To 10 kHz |
| Maximum output compliance voltage |
4.5 Vrms (6.4 Vpeak) |
| Inductive drive capability: |
1 mH load |
| Output isolation, current terminal to earth |
600 Vrms, 850 Vpeak, dc to 850 Hz |
| Performance specifications |
| AC accuracy, closed loop with 6105A |
To 140 ppm |
| DC accuracy, standalone |
To 160 ppm |
| AC accuracy, standalone |
To 350 ppm |
| Phase angle accuracy |
0.006° |
| Load dependent phase angle shift |
<0.001° @ 60 Hz |
| General specifications |
| Input line voltage range |
100 V to 240 V with up to ±10 % fluctuations |
| Transient overvoltage |
Impulse withstand (overvoltage); Category II of IEC 60364-4-443 |
| Frequency |
47 Hz to 63 Hz |
| Maximum consumption |
< 1500 VA |
| Calibration Documentation |
Report of calibration with data included; 17025 accredited report optional |
| Dimensions with feet (H x W x L) |
192 mm x 432 mm x 645 mm
(7.6 inches x 17 inches x 25.5 inches) |
| Dimensions without feet (H x W x L) |
178 mm x 432 mm x 645 mm
(7 inches x 17 inches x 25.5 inches) |
| Weight |
25 kg (54 lb.) |
| Design standards and compliance |
Designed to EN 61010-1: 2010, CAN/CSA 22.2 No 61010.1-04, ANSI/UL 61010-1:2004, EN 61326-1:2006 CE marked, CSA listed |